VASCUTRAK PTA DILATATION CATHETER V184040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-01-15 for VASCUTRAK PTA DILATATION CATHETER V184040 manufactured by Bard Peripheral Vascular, Inc..

Event Text Entries

[179403892] The lot number for the malfunction was provided and a lot history review was performed. The device was returned for evaluation; the investigation was confirmed for the reported balloon rupture, as a longitudinal rupture was noted to the balloon. The investigation was also confirmed for core wire break, as a break was noted to core wire. Based upon the available information, the definitive root cause is unknown. The device is labeled for single use.
Patient Sequence No: 1, Text Type: N, H10


[179403893] This report summarizes one malfunction. A review of the reported information indicated that model v184040 vascutrak pta dilatation catheter allegedly experienced balloon longitudinal material rupture and break. This information was received from one source. Of the one balloon longitudinal material rupture malfunction, one patient was involved with no patient consequence or impact. Patient's age, weight, and gender was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020394-2020-00422
MDR Report Key9594168
Date Received2020-01-15
Date of Report2020-01-15
Date Mfgr Received2019-12-31
Date Added to Maude2020-01-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJUDITH LUDWIG
Manufacturer Street1415 W. 3RD STREET
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal85281
Manufacturer Phone4803032689
Manufacturer G1C.R. BARD, INC. (GFO)
Manufacturer Street289 BAY ROAD
Manufacturer CityQUEENSBURY NY 12804
Manufacturer CountryUS
Manufacturer Postal Code12804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVASCUTRAK PTA DILATATION CATHETER
Generic NamePTA BALLOON DILATATION CATHETER
Product CodePNO
Date Received2020-01-15
Model NumberV184040
Catalog NumberV184040
Lot NumberGFDR4288
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBARD PERIPHERAL VASCULAR, INC.
Manufacturer Address1625 W 3RD ST. TEMPE AZ 85281 US 85281


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-15

© 2026 FDA.report