ACHIEVA 1.5T NEW 781296 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-20 for ACHIEVA 1.5T NEW 781296 NA manufactured by Philips Healthcare.

Event Text Entries

[175737172] Based on the provided information and tests performed on site, there is no indication of a malfunction of the mri system or coil used that contributed to the blistering. The injury is concluded to have been caused by a conductive material within the patients? Clothing. Contributing factors in this case: 10 scans with high sar values (> 2 w/kg) were executed in a short period of time, allowing no cool down time for the patient. The total administered specific energy dose of 4. 8 kj/kg exceeded the recommended limit of 3. 5 kj/kg.
Patient Sequence No: 1, Text Type: N, H10


[175737173] Philips received a report related to a heating incident (striped shaped blistering on the patient? S abdomen) on an achieva 1. 5t mr system.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003768277-2020-00005
MDR Report Key9606869
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-20
Date of Report2019-12-27
Date of Event2019-12-19
Date Mfgr Received2019-12-27
Device Manufacturer Date2012-05-16
Date Added to Maude2020-01-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DUSTY LEPPERT
Manufacturer StreetVEENPLUIS 4-6 P.O. BOX 10.000
Manufacturer CityBEST 5680DA
Manufacturer CountryNL
Manufacturer Postal5680 DA
Manufacturer G1PHILIPS MEDICAL SYSTEMS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal Code01810
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACHIEVA 1.5T NEW
Generic NameSYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Product CodeLNH
Date Received2020-01-20
Model Number781296
Catalog NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS HEALTHCARE
Manufacturer AddressVEENPLUIS 4-6 P.O. BOX 10.000 BEST 5680DA NL 5680 DA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.