AIRSEAL 12MM ACCESS PORT AND PALM GRIP OBTURATOR WITH BLADELESS OPTICAL TIP IAS12-100LPI

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-01-20 for AIRSEAL 12MM ACCESS PORT AND PALM GRIP OBTURATOR WITH BLADELESS OPTICAL TIP IAS12-100LPI manufactured by Sequel Special Products.

Event Text Entries

[181401311] At time of filing, although expected, the reported device has not been returned to conmed for evaluation. A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation. This issue will continue to be monitored through the complaint system to assure patient safety.
Patient Sequence No: 1, Text Type: N, H10


[181401312] This record captures the issue with the ias12-100lpi, airseal 12mm access port and palm grip obturator with bladeless optical tip, 100mm length, lot 99419 as reported by the facility to conmed (b)(4). It is reported that on (b)(6) 2019, during a "ralp" procedure the trocar cap rubber valve broke, and the fragment fell into the abdominal cavity. The fragments have been collected immediately. It is indicated that the procedure was successfully completed. Additional information received confirmed the procedure was a robot-assisted laparoscopic prostatectomy. While the surgeon was manually operating surgical instruments through the trocar, the rubber valve fell off in 1 piece into the patient. The piece was retrieved through the trocar by using forceps. It was confirmed that there was no impact or injury to the patient and the procedure was completed as scheduled & planned. This report is being raised on the basis of previous us fda mdr reporting of similar malfunction with potential for injury upon reoccurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1320894-2019-00452
MDR Report Key9608264
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-01-20
Date of Report2020-02-07
Date of Event2019-12-23
Date Mfgr Received2020-02-06
Date Added to Maude2020-01-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS TRACEY WEISELBENTON
Manufacturer Street11311 CONCEPT BLVD
Manufacturer CityLARGO FL 337734908
Manufacturer CountryUS
Manufacturer Postal337734908
Manufacturer Phone7273995557
Manufacturer G1SEQUEL SPECIAL PRODUCTS
Manufacturer Street122 AVENUE OF INDUSTRY PO BOX 11050
Manufacturer CityWATERBURY CT 067035994
Manufacturer CountryUS
Manufacturer Postal Code067035994
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAIRSEAL 12MM ACCESS PORT AND PALM GRIP OBTURATOR WITH BLADELESS OPTICAL TIP
Generic NameINSUFFLATOR, LAPAROSCOPIC
Product CodeHIF
Date Received2020-01-20
Returned To Mfg2020-01-15
Catalog NumberIAS12-100LPI
Lot Number99419
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSEQUEL SPECIAL PRODUCTS
Manufacturer Address122 AVENUE OF INDUSTRY PO BOX 11050 WATERBURY CT 06703 US 06703


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-20

© 2026 FDA.report