TORNADO FD1810

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-01-22 for TORNADO FD1810 manufactured by Getinge Ic Production Poland Sp. Z O.o.

Event Text Entries

[175835395] Please note that previous medwatch reports for this product may have been submitted under the following registration numbers: (b)(4). Currently, these products are to be handled by arjohuntleigh ab? S complaint handling establishment and any medwatch reports will be submitted under registration #(b)(4). The involved device was evaluated by the arjo representative and found unusable, hence it was removed from service. The investigations on-going and further information will be provided in the next report.
Patient Sequence No: 1, Text Type: N, H10


[175835396] Arjo was notified about an event with involvement of tornado flusher. It was reported that the device's steam generator started the fire, which burned the quick connect cable and all components around that part. No other part was damaged. The facility personnel used the extinguisher to stop the fire.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420694-2020-00009
MDR Report Key9615562
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-01-22
Date of Report2020-03-19
Date of Event2019-12-26
Date Mfgr Received2019-12-27
Device Manufacturer Date2015-06-23
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKINGA STOLINSKA
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI 62-052
Manufacturer CountryPL
Manufacturer Postal62-052
Manufacturer Phone688282467
Manufacturer G1ARJOHUNTLEIGH MAGOG INC.
Manufacturer Street2001 TANGUAY STREET
Manufacturer CityMAGOG J1X5Y5
Manufacturer CountryCA
Manufacturer Postal CodeJ1X5Y5
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTORNADO
Generic NameWASHER OF BODY WASTE RECEPTACLES
Product CodeFLH
Date Received2020-01-22
Model NumberFD1810
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGETINGE IC PRODUCTION POLAND SP. Z O.O
Manufacturer AddressUL. SZKOLNA 30, PLEWISKA PLEWISKA PL-62064 PL PL-62064


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.