SOMATOM DEFINITION AS 8098027

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-22 for SOMATOM DEFINITION AS 8098027 manufactured by Siemens Healthcare Gmbh.

Event Text Entries

[175847023] Initial reporter's telephone number is: (b)(6). Siemens has completed a technical investigation of the reported event. The analysis of the siemens ct logfiles do not show a malfunction of siemens-manufactured portion of the system. The siemens ct system performed as specified. The log file review also revealed that the non-siemens injector did not generate an error, hence the reason why there was no error message in the ct logfile. Further siemens investigation is not warranted at this time.
Patient Sequence No: 1, Text Type: N, H10


[175847024] It was reported to siemens that a patient passed away during a ct contrast study on (b)(6) 2020 at approximately 5:37am. According to the initial investigation, the pistol for the contrast medium supply of the injector (a non-siemens device) jammed and due to this issue, air was added to the contrast medium. The patient subsequently died due to an air embolism. Since the siemens ct system has an interface to the injector device, this event is being reported by siemens. The reported event occurred in (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004977335-2020-13847
MDR Report Key9615750
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-22
Date of Report2020-01-02
Date of Event2020-01-01
Date Mfgr Received2020-01-02
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactREBECCA TUDOR
Manufacturer Street40 LIBERTY BLVD. 65-1A
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6104486484
Manufacturer G1SIEMENS HEALTHCARE GMBH
Manufacturer StreetSIEMENSTRASSE 1
Manufacturer CityFORCHHEIM, 91301
Manufacturer CountryGM
Manufacturer Postal Code91301
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSOMATOM DEFINITION AS
Generic NameSYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Product CodeJAK
Date Received2020-01-22
Model Number8098027
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE GMBH
Manufacturer AddressSIEMENSTRASSE 1 FORCHHEIM, 91301 GM 91301


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-01-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.