MAIN CART 9735669 STEALTH S8 EM ENT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-22 for MAIN CART 9735669 STEALTH S8 EM ENT manufactured by Medtronic Navigation, Inc.

Event Text Entries

[183931654] Other relevant device(s) are: product id: 9735736, serial/lot #: (b)(4). A medtronic representative went to the site to test the equipment. It was reported that no parts of the navigation system was replaced. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[183931655] Medtronic received information regarding a navigation system that was used during a functional endoscopic sinus surgery (fess) procedure. It was reported that during the case, an error 64 show up while the surgeon was on the registration screen. The surgeon aborted navigation for the case, but a site representative was able to clear the error message with a reboot and load exams up for cases and view them on the registration screen. There was reported delay of less than one hour due to the issue. There is no known impact on patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1723170-2020-00214
MDR Report Key9616724
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-22
Date of Report2020-01-22
Date of Event2020-01-20
Date Mfgr Received2020-01-20
Device Manufacturer Date2018-04-11
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAIN CART 9735669 STEALTH S8 EM ENT
Generic NameEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Product CodePGW
Date Received2020-01-22
Model Number9735669
Catalog Number9735669
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.