MINDRAY MPM MULTI PARAMETER MODULE FMMR1000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-01-22 for MINDRAY MPM MULTI PARAMETER MODULE FMMR1000 manufactured by Shenzhen Mindray Bio-meidcal Electronics Co., Ltd.

Event Text Entries

[181554404] The complaint device was returned for evaluation. Visual inspection found no anomalies. Device evaluation identified an spo2 malfunction confirming the reported event. The spo2 connector was refurbished. The front and rear cases were inspected. Diagnostics were ran and all pcbs were tested. The case was checked for damage. The root cause for the reported issue was determined to be a bad solder joint. This type of reported event will continue to be monitored. It should be noted that this is being filed based on retrospective review.
Patient Sequence No: 1, Text Type: N, H10


[181554405] Reportedly, post repair, the device was still having intermittent spo2 issues. There was no report of patient harm or patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007409280-2020-00011
MDR Report Key9616841
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2020-01-22
Date of Report2018-07-18
Date Mfgr Received2018-07-18
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNA VARGAS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal92672
Manufacturer Phone8004495328
Manufacturer G1AVANTE HEALTH SOLUTIONS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal Code92672
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMINDRAY MPM MULTI PARAMETER MODULE
Generic NameHYPERVISOR CENTRAL MONITORING SYSTEM
Product CodeMSX
Date Received2020-01-22
Returned To Mfg2018-07-23
Model NumberFMMR1000
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSHENZHEN MINDRAY BIO-MEIDCAL ELECTRONICS CO., LTD
Manufacturer Address800 MACARTHUR BLVD. MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-22

© 2026 FDA.report