REMISOL ADVANCE (RADV) WORK STATION, US HW, RADV B25255

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-01-22 for REMISOL ADVANCE (RADV) WORK STATION, US HW, RADV B25255 manufactured by Normand-info S.a.s.u..

Event Text Entries

[176281818] Beckman coulter customer technical service (cts) evaluated the event, and identified the "validation range" settings were incorrect. The patient was one (1) day old and was not included in the "less that 1 day old" validation range. Cts confirmed that the rules ran as expected based on the "validation range" settings. The settings were adjusted to resolve the issue. Age or date of birth, weight, and ethnicity: information not provided by customer. (b)(6). The beckman coulter internal identifier is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[176281819] The customer reported critical tbil (total bilirubin) results for an infant were auto validated by remisol advance (radv) when it should have not been auto validated, which caused a delay in patient treatment by one (1) day. There was no report of injury or death associated with this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006543086-2020-00001
MDR Report Key9617299
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-01-22
Date of Report2020-01-22
Date of Event2020-01-03
Date Facility Aware2020-01-03
Date Mfgr Received2020-01-03
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. CHRISTINE XARDEL
Manufacturer Street7 RUE FREDERIC DEGEORGE BAT B 2 IEME ETAGE
Manufacturer CityARRAS, 62000
Manufacturer CountryFR
Manufacturer Postal62000
Manufacturer G1NORMAND-INFO S.A.S.U.
Manufacturer Street7 RUE FREDERIC DEGEORGE BAT B 2IEME ETAGE
Manufacturer CityARRAS PAS-DE-CALAIS, 62028
Manufacturer CountryFR
Manufacturer Postal Code62028
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameREMISOL ADVANCE (RADV)
Generic NameCALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE
Product CodeJQP
Date Received2020-01-22
Model NumberWORK STATION, US HW, RADV
Catalog NumberB25255
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNORMAND-INFO S.A.S.U.
Manufacturer Address7 RUE FREDERIC DEGEORGE BAT B 2 IEME ETAGE ARRAS, 62000 FR 62000


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-22

© 2024 FDA.report