SAMSUNG SAMSUNG TAB E SM-T377A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-01-22 for SAMSUNG SAMSUNG TAB E SM-T377A manufactured by Visicu, Inc - Dup.

Event Text Entries

[180760295] Investigation and objective evidence revealed no burning, melting or charring of the adaptor or cord. The adaptor was identified as an unapproved/supported/supplied adaptor used by the complainant.
Patient Sequence No: 1, Text Type: N, H10


[180760296] It was reported on (b)(6) 2019 that the cord to the tablet was on fire. Nurse asked patient if there were active sparks or fire. Patient reported she was unsure. Nurse confirmed there was no injury to the patient. Installer returned to the patient home on (b)(6) 2020 to retrieve the cord and tablet and subsequently returned the entire kit including the ecad bp, spo2, and the scale which is the entire kit.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1125873-2020-00001
MDR Report Key9617491
Report SourceHEALTH PROFESSIONAL
Date Received2020-01-22
Date of Report2020-01-02
Date Mfgr Received2019-12-31
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SHARON FISHTER
Manufacturer Street217 E REDWOOD ST STE 1900
Manufacturer CityBALTIMORE MD 21202
Manufacturer CountryUS
Manufacturer Postal21202
Manufacturer G1PHILIPS HEALTHCARE (ALPHARETTA)
Manufacturer StreetONE DEERFIELD CENTER 13560 MORRIS ROAD STE 2100
Manufacturer CityALPHARETTA GA 30004
Manufacturer CountryUS
Manufacturer Postal Code30004
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSAMSUNG
Generic NameN/A
Product CodeMSX
Date Received2020-01-22
Model NumberSAMSUNG TAB E SM-T377A
Lot NumberNA
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVISICU, INC - DUP
Manufacturer Address217 E REDWOOD ST STE 1900 BALTIMORE MD 21202 US 21202


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.