AMS SPECTRA CONCEALABLE PENILE PROSTHESIS 720074-03

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-01-22 for AMS SPECTRA CONCEALABLE PENILE PROSTHESIS 720074-03 manufactured by Boston Scientific Corporation.

Event Text Entries

[180581050] It was reported that after experiencing infection and proximal erosion with another company's penile prosthesis the patient underwent a surgical procedure in which a new spectra penile prosthesis (spp) was implanted. However, on a follow up appointment the physician noted the patient was experiencing distal erosion of one cylinder. It was mentioned that this did not come as a surprise as the patient had very poor tissue quality. A surgical procedure was performed in which both spp cylinders were removed. The patient did not experience any further complications. It was further reported that the erosion was identified due to the implant coming out of the patient's glans. However, physician does not believe the device caused the erosion. It was indicated that the previous company{s device likely caused pending erosion due to patient leaving it inflated for six years resulting in poor tissue quality. There were no device malfunctions associated with the explanted spp.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183959-2020-00208
MDR Report Key9618130
Report SourceHEALTH PROFESSIONAL
Date Received2020-01-22
Date of Report2020-01-22
Date of Event2020-01-07
Date Mfgr Received2020-01-08
Device Manufacturer Date2017-12-12
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALYSON HARRIS
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal55343
Manufacturer Phone4083953452
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal Code55343
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Generic NamePROSTHESIS PENILE
Product CodeFAE
Date Received2020-01-22
Model Number720074-03
Catalog Number720074-03
Lot Number1000023096
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address10700 BREN ROAD W MINNETONKA MN 55343 US 55343


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-01-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.