INOMAX DSIR (DELIVERY SYSTEM) 10071

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-01-22 for INOMAX DSIR (DELIVERY SYSTEM) 10071 manufactured by Mallinckrodt Manufacturing Llc.

Event Text Entries

[181373893] Reportable malfunction with inomax dsir dg20150119 was documented and investigated under (b)(4). A review of the device's service log revealed that a delivery failure alarm occurred due to over-delivery of nitric oxide for 12 consecutive seconds on 16-aug-2019. The root cause for this occurrence was likely due to a malfunctioning proportional valve. As a result, it was recommended to the distributor to replace the proportional valve. This condition will be tracked and trended under the company's quality system. Trends were reviewed for this reportable condition and determined to be within the established control limits. No further action is required at this time.
Patient Sequence No: 1, Text Type: N, H10


[181373894] On (b)(6) 2019, a mallinckrodt internal employee discovered a delivery failure alarm due overdelivery for inomax dsir dg20150119. This service log finding meets the criteria of a reportable malfunction. There was no complaint alleged against this device. This match was found during routine review of service logs from a device in the (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004531588-2020-00016
MDR Report Key9618235
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-01-22
Date of Report2020-01-22
Date of Event2019-08-16
Date Mfgr Received2019-12-23
Device Manufacturer Date2015-05-14
Date Added to Maude2020-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMEGAN VERNAK
Manufacturer Street1425 US ROUTE 206
Manufacturer CityBEDMINSTER NJ 07921
Manufacturer CountryUS
Manufacturer Postal07921
Manufacturer G1MALLINCKRODT MANUFACTURING LLC
Manufacturer Street6603 FEMRITE DRIVE
Manufacturer CityMADISON WI 53718
Manufacturer CountryUS
Manufacturer Postal Code53718
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINOMAX DSIR (DELIVERY SYSTEM)
Generic NameAPPARATUS
Product CodeMRN
Date Received2020-01-22
Model Number10071
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMALLINCKRODT MANUFACTURING LLC
Manufacturer Address6603 FEMRITE DRIVE MADISON WI 53718 US 53718


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.