TRAPEZOID RX M00510860 1086

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-24 for TRAPEZOID RX M00510860 1086 manufactured by Boston Scientific Corporation.

Event Text Entries

[176247176] (b)(4). The device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[176247177] It was reported to boston scientific corporation that a trapezoid rx basket was used in the biliary duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019. According to the complainant, during the procedure, an alliance handle was used in conjunction with the trapezoid basket in an attempt to crush a stone. However, the handle cannula broke during the attempt, entrapping the stone inside the basket and causing the basket to become stuck inside the patient's biliary duct. The catheter was cut at the basket handle and the scope was removed from the patient. The stone and basket remained in the bile duct and the patient was referred to another hospital. Reportedly, the condition of the patient at the conclusion of the procedure was stable, however, a few days after a bile drain tube insertion, the patient passed away. The basket remained in the patient at the time of the patient's death.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2020-00107
MDR Report Key9627350
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-24
Date of Report2020-02-18
Date of Event2019-12-27
Date Mfgr Received2020-01-24
Device Manufacturer Date2019-04-12
Date Added to Maude2020-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRAPEZOID RX
Generic NameLITHOTRIPTOR, BILIARY MECHANICAL
Product CodeLQC
Date Received2020-01-24
Model NumberM00510860
Catalog Number1086
Lot Number0023638740
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-01-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.