CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-01-24 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[176381295] Product analysis : analysis confirmed original complaint. The behavior is linked to a known issue, rpi 108712. The interrogation hangs when initiated while the application is in background. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[176381296] It was reported that during use of an electrocautery device, the programmer had been placed on the cart next to the cautery machine. When the physician started using cautery, the programmer displayed a dotted line for the connection status. When cautery was turned off, the connection resumed. When the physician continued use of cautery, the application completely crashed and the home screen was displayed. In addition, when first pulling up the programmer, and trying to pair the patient connector, the application completely froze. A spinning hour glass displayed, and the application was frozen for five minutes. The user closed the application to restart. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004593495-2020-00061
MDR Report Key9629918
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-01-24
Date of Report2020-01-24
Date of Event2019-10-08
Date Mfgr Received2020-01-21
Device Manufacturer Date2019-03-22
Date Added to Maude2020-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-01-24
Model Number24970A
Catalog Number24970A
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-24

© 2026 FDA.report