CURVED OSTEOTOME N/A 00705300301

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2020-01-27 for CURVED OSTEOTOME N/A 00705300301 manufactured by Zimmer Tmt.

Event Text Entries

[176658133] (b)(4). Complaint sample was evaluated and the reported event was confirmed. Visual examination of the returned device identified that the clamping screw was no longer functioning as intended, as it would not tighten. The hex threads inside the screw were damaged likely from tightening the screw. The device had an approximate field life of 3 years. Device history record (dhr) was reviewed and no discrepancies were found. Root cause was unable to be determined. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends. Report source: event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[176658154] It was reported that during a primary tha, the instrument would not hold the blade component. The surgeon was unable to use the device and proceeded to use a chisel for surgery. This resulted in a 31-60 minute delay in surgery. Attempts have been made and no further information has been provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005751028-2020-00008
MDR Report Key9631457
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2020-01-27
Date of Report2020-01-27
Date of Event2019-10-01
Date Mfgr Received2020-01-23
Device Manufacturer Date2016-06-10
Date Added to Maude2020-01-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM HARA
Manufacturer Street10 POMEROY ROAD
Manufacturer CityPARSIPPANY NJ 07054
Manufacturer CountryUS
Manufacturer Postal07054
Manufacturer Phone9735760032
Manufacturer G1ZIMMER TMT
Manufacturer Street10 POMEROY ROAD
Manufacturer CityPARSIPPANY NJ 07054
Manufacturer CountryUS
Manufacturer Postal Code07054
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCURVED OSTEOTOME
Generic NameINSTRUMENT, SURGICAL
Product CodeHWM
Date Received2020-01-27
Returned To Mfg2019-11-18
Model NumberN/A
Catalog Number00705300301
Lot Number63396488
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER TMT
Manufacturer Address10 POMEROY ROAD PARSIPPANY NJ 07054 US 07054


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2020-01-27

© 2024 FDA.report