[176465421] Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
Patient Sequence No: 1, Text Type: N, H10


[176465422] Complainant alleged that during functional testing, these internal handles failed to discharge. Complainant did not indicate that there was any patient involvement in the reported malfunction.
Patient Sequence No: 1, Text Type: D, B5


[186067368] The internal handles were returned to zoll medical corporation and the customer's report was not replicated or confirmed. The internal handles passed testing and were able to discharge at all energy levels with no faults found. The internal handles were forwarded to scrap as a precaution. Analysis of reports of this type has not identified an increase in trend.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1220908-2020-00079
MDR Report Key9633260
Report SourceFOREIGN
Date Received2020-01-27
Date of Report2020-01-06
Date Mfgr Received2020-01-06
Device Manufacturer Date2018-03-01
Date Added to Maude2020-01-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street269 MILL ROAD
Manufacturer CityCHELMSFORD MA 01824
Manufacturer CountryUS
Manufacturer Postal01824
Manufacturer Phone9784219552
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAUTOCLAVABLE INTERNAL HANDLES (MSERIES)
Generic NameINTERNAL PADDLE
Product CodeLDD
Date Received2020-01-27
Returned To Mfg2020-01-27
Model Number8011-0501-01
Catalog Number8011050101
Lot NumberNA
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL MEDICAL CORPORATION
Manufacturer Address269 MILL ROAD CHELMSFORD MA 01824 US 01824


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-27