MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-27 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.
[176499145]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[176499146]
It was reported that after updating the programmer, difficulties were experienced when programming an implantable device during implant. The "lead implant date" and "device implant date" were noted to not turn blue after changing them and after pressing "program", they were noted to remain blank. It was noted that the implantable device had been successfully programmed with a programmer that had not yet been updated. The same issue was noted to have occurred on multiple instances with different tablets. Troubleshooting was performed and it was noted that the update had changed the behavior of native controls for the programmer. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
[183791761]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[188584841]
Product event summary: the behavior of the native controls (date picker, keyboard) has changed to where-in - one tap outside of the control is needed to close the control (date picker, keyboard) and another tap is needed to reflect the changed value. As such, for setting a date, if you tap outside the date picker twice, you see the changed value shown as 'pending'. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3004593495-2020-00066 |
MDR Report Key | 9634150 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-01-27 |
Date of Report | 2020-03-02 |
Date of Event | 2020-01-01 |
Date Mfgr Received | 2020-02-26 |
Date Added to Maude | 2020-01-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Street | BAYAN LEPAS FREE INDUSTRIAL ZO |
Manufacturer City | BAYAN LEPAS 11900 |
Manufacturer Country | MY |
Manufacturer Postal Code | 11900 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CARELINK SMARTSYNC BASE |
Generic Name | ANALYZER, PACEMAKER GENERATOR FUNCTION |
Product Code | DTC |
Date Received | 2020-01-27 |
Model Number | 24970A |
Catalog Number | 24970A |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Address | BAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-01-27 |