CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-27 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[176499145] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[176499146] It was reported that after updating the programmer, difficulties were experienced when programming an implantable device during implant. The "lead implant date" and "device implant date" were noted to not turn blue after changing them and after pressing "program", they were noted to remain blank. It was noted that the implantable device had been successfully programmed with a programmer that had not yet been updated. The same issue was noted to have occurred on multiple instances with different tablets. Troubleshooting was performed and it was noted that the update had changed the behavior of native controls for the programmer. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[183791761] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[188584841] Product event summary: the behavior of the native controls (date picker, keyboard) has changed to where-in - one tap outside of the control is needed to close the control (date picker, keyboard) and another tap is needed to reflect the changed value. As such, for setting a date, if you tap outside the date picker twice, you see the changed value shown as 'pending'. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004593495-2020-00066
MDR Report Key9634150
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-27
Date of Report2020-03-02
Date of Event2020-01-01
Date Mfgr Received2020-02-26
Date Added to Maude2020-01-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-01-27
Model Number24970A
Catalog Number24970A
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-27

© 2026 FDA.report