3 LESION NT1100? PAIN MANAGEMENT RF GENERATOR RFG-NT-1100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-28 for 3 LESION NT1100? PAIN MANAGEMENT RF GENERATOR RFG-NT-1100 manufactured by Neurotherm, Inc.

Event Text Entries

[181618527] The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported sparking was attributed to the use environment. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. The review determined the process was performed and completed in accordance with abbott specifications and procedures.
Patient Sequence No: 1, Text Type: N, H10


[181618528] The generator had water damage due to flooding and ports 2 and 3 would not function. During functional testing, the impedance in port 2 was 1400 ohms and port 3 would not display impedance. Additionally, an "unsafe probe" error message occurred for both ports. When testing the lesion, visible sparking was observed on the grounding pad at the site of probe 3. There was no patient involved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002953813-2020-00005
MDR Report Key9636276
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-28
Date of Report2020-01-28
Date of Event2020-01-15
Date Mfgr Received2020-01-15
Device Manufacturer Date2010-05-05
Date Added to Maude2020-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1NEUROTHERM, INC
Manufacturer Street600 RESEARCH DRIVE STE 1
Manufacturer CityWILMINGTON MA 01887
Manufacturer CountryUS
Manufacturer Postal Code01887
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3 LESION NT1100? PAIN MANAGEMENT RF GENERATOR
Generic NameGENERATOR, LESION, RADIOFREQUENCY
Product CodeGXD
Date Received2020-01-28
Model NumberRFG-NT-1100
Catalog NumberRFG-NT-1100
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNEUROTHERM, INC
Manufacturer Address600 RESEARCH DRIVE STE 1 WILMINGTON MA 01887 US 01887


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.