4 LESION NT2000? PAIN MANAGEMENT RF GENERATOR RFG-NT-2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-28 for 4 LESION NT2000? PAIN MANAGEMENT RF GENERATOR RFG-NT-2000 manufactured by St. Jude Medical, Inc..

Event Text Entries

[176647472] One 4 lesion nt2000 rf generator was received for analysis. Visual inspection of the returned generator revealed burnt components on the upper assembly. Ac power was applied to the rf generator and the system powered on with a continuous beep confirming the reported event. The upper assembly was replaced with a good upper assembly and the issue was resolved. Based on the information provided to abbott and the investigation performed, the cause for the reported event was due to an abnormally functioning generator upper assembly.
Patient Sequence No: 1, Text Type: N, H10


[176647473] This event is being reported based on analysis which revealed burnt components in the generator.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2184149-2020-00011
MDR Report Key9638610
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-28
Date of Report2020-01-28
Date of Event2019-12-10
Date Mfgr Received2020-01-22
Device Manufacturer Date2015-09-16
Date Added to Maude2020-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, INC.
Manufacturer StreetONE ST. JUDE MEDICAL DRIVE
Manufacturer CityST. PAUL MN 55117
Manufacturer CountryUS
Manufacturer Postal Code55117
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name4 LESION NT2000? PAIN MANAGEMENT RF GENERATOR
Generic NameGENERATOR, LESION, RADIOFREQUENCY
Product CodeGXD
Date Received2020-01-28
Model NumberRFG-NT-2000
Catalog NumberRFG-NT-2000
Lot Number6028094
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.
Manufacturer AddressONE ST. JUDE MEDICAL DRIVE ST. PAUL MN 55117 US 55117


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.