MIDWEST TRADITION HANDPIECE 780045

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-01-28 for MIDWEST TRADITION HANDPIECE 780045 manufactured by Sirona Dental Systems Gmbh.

Event Text Entries

[184711070] Additional serial numbers included in this report: (b)(4). 9 of 10 devices were returned for evaluation. 1 device will not be returned for evaluation. Evaluation of 3 devices found chuck wear and lack of maintenance. The devices were cleaned of debris and the turbines were replaced. The devices were repaired and returned to the customer. Evaluation of 6 devices found normal chuck wear and the turbines needed to be replaced. The devices were repaired and returned to the customers.
Patient Sequence No: 1, Text Type: N, H10


[184711071] This report summarizes 10 malfunction events. This report summarizes ten malfunction events where a midwest tradition handpiece won't hold burs. There were no injuries in any of the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9614977-2019-00169
MDR Report Key9639382
Date Received2020-01-28
Date of Report2020-01-28
Date Added to Maude2020-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494424
Manufacturer G1SIRONA DENTAL SYSTEMS GMBH
Manufacturer StreetFABRIKSTRASSE 31
Manufacturer CityBENSHEIM, D-64625
Manufacturer CountryGM
Manufacturer Postal CodeD-64625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMIDWEST TRADITION HANDPIECE
Generic NameHANDPIECE, AIR-POWERED, DENTAL
Product CodeEFB
Date Received2020-01-28
Model NumberNA
Catalog Number780045
Lot NumberNA
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerSIRONA DENTAL SYSTEMS GMBH
Manufacturer AddressFABRIKSTRASSE 31 BENSHEIM, D-64625 GM D-64625


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.