PNSA STARTER BP EN ES USA 101035077

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-01-28 for PNSA STARTER BP EN ES USA 101035077 manufactured by Medela Llc.

Event Text Entries

[176751185] The customer was sent replacement 21mm personal fit breast shields so that she could return them to a retail location for a refund and her original breast shields were requested to be returned for testing/evaluation. The customer was contacted by a complaint handler on multiple occasions, including in writing, to get additional information, with no response as of the date of this report. Medela is filing this report, which is considered a serious injury as it required medical attention (medication was prescribed). Reported issues of rash were investigated under (b)(4), which determined the root cause to be potential for the product, after it is opened, to come in contact with environmental allergens, lotions or creams, and cleaning solutions that may cause an allergic reaction or irritation. Such root cause is not a product malfunction or a deficiency in information available to the user.
Patient Sequence No: 1, Text Type: N, H10


[176751186] On (b)(6) 2020, the customer alleged to medela llc that the 21mm personal fit breast shields she was using with her pump in style breast pump had caused her to develop a rash. She additionally alleged that she had seen a dermatologist and her obgyn doctor, who advised her that it was an allergic reaction to polypropylene and she was prescribed a steroid as treatment for two weeks.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419937-2020-00005
MDR Report Key9639563
Report SourceCONSUMER
Date Received2020-01-28
Date of Event2020-01-07
Date Mfgr Received2020-01-07
Date Added to Maude2020-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAN KLOIBER
Manufacturer Street1101 CORPORATE DRIVE
Manufacturer CityMCHENRY IL 60050
Manufacturer CountryUS
Manufacturer Postal60050
Manufacturer Phone8004358316
Manufacturer G1MEDELA LLC
Manufacturer Street1101 CORPORATE DRIVE
Manufacturer CityMCHENRY IL 60050
Manufacturer CountryUS
Manufacturer Postal Code60050
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePNSA STARTER BP EN ES USA
Generic NamePUMP, BREAST, POWERED
Product CodeHGX
Date Received2020-01-28
Model Number101035077
Catalog Number101035077
Lot NumberNI
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDELA LLC
Manufacturer Address1101 CORPORATE DRIVE MCHENRY IL 60050 US 60050


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-28

© 2026 FDA.report