STRETCHAIR PT TRANSFER SYSTEMS, INC. MC675CLTP UMK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2006-04-27 for STRETCHAIR PT TRANSFER SYSTEMS, INC. MC675CLTP UMK manufactured by Stretchair Patient Transfer Systems, Inc..

Event Text Entries

[18831621] Pt moved himself up in chair with assist from nursing staff. Later he stated he leaned forward and the buckle on the seatbelt came loose and he fell forward out of chair.
Patient Sequence No: 1, Text Type: D, B5


[18907889] Eval of the seat belt indicated under normal conditions that functioned as required. When the seat belt received a significant dynamic pull on each side of the latch, the latch did release. The internal spring within the latching mechanism was not properly engaging the locking clip. All current inventories of seat belts have been checked and no defective units have been found. This is an isolated case.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1043565-2005-00001
MDR Report Key964107
Report Source06
Date Received2006-04-27
Date of Report2006-03-28
Date of Event2006-03-01
Date Facility Aware2006-03-17
Report Date2006-03-31
Date Reported to FDA2006-03-31
Date Reported to Mfgr2006-03-31
Date Mfgr Received2006-04-04
Device Manufacturer Date2006-01-01
Date Added to Maude2007-12-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street8110 ULMERTON RD
Manufacturer CityLARGO FL 33771
Manufacturer CountryUS
Manufacturer Postal33771
Manufacturer Phone7275312444
Manufacturer G1STRETCHAIR PTS, INC.
Manufacturer Street8110 ULMERTON RD
Manufacturer CityLARGO FL 33771
Manufacturer CountryUS
Manufacturer Postal Code33771
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRETCHAIR PT TRANSFER SYSTEMS, INC.
Generic NameSTRETCHAIR
Product CodeIQC
Date Received2006-04-27
Model NumberMC675CLTP
Catalog NumberUMK
Lot NumberUNK
ID NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeUNKNOWN
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key684526
ManufacturerSTRETCHAIR PATIENT TRANSFER SYSTEMS, INC.
Manufacturer Address8110 ULMERTON RD. LARGO FL 33771 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-04-27