WAYNE PNEUMOTHORAX TRAY C-UTPTY-1400-WAYNE-112497-IMH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other,use report with the FDA on 2020-01-29 for WAYNE PNEUMOTHORAX TRAY C-UTPTY-1400-WAYNE-112497-IMH manufactured by Cook Inc.

Event Text Entries

[177369546] Pma/510(k) #: exempt. This report includes information known at this time. A follow-up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[177369547] It was reported that the guidewire in a wayne pneumothorax tray was found to be broken upon removal. The product failed upon initial placement during a thoracostomy procedure to drain a hemothorax. The device was placed in the right anterior chest, 4th intercostal space, to drain bodily fluid. The catheter was placed and resistance was felt upon removing the guidewire. When pulled out, it was found to be broken. After the guide wire was observed to be broken, "the straightening trocar was removed". The catheter "never drained blood", so a second device was successfully used in its place. No adverse effects to the patient have been reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00233
MDR Report Key9643631
Report SourceHEALTH PROFESSIONAL,OTHER,USE
Date Received2020-01-29
Date of Report2020-03-19
Date of Event2020-01-13
Date Mfgr Received2020-03-12
Device Manufacturer Date2019-12-04
Date Added to Maude2020-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAYNE PNEUMOTHORAX TRAY
Generic NameJOL CATHETER AND TIP, SUCTION
Product CodeJOL
Date Received2020-01-29
Model NumberNA
Catalog NumberC-UTPTY-1400-WAYNE-112497-IMH
Lot Number10158554
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-29

© 2026 FDA.report