CELL SAVER ELITE PROCESSING SET (225 ML) CSE-P-225

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-29 for CELL SAVER ELITE PROCESSING SET (225 ML) CSE-P-225 manufactured by Haemonetics Corporation.

Event Text Entries

[176869035] The field service engineer went onsite to gather data on the procedure and discovered the leak occurred on cycle 4, with the total volume reinfused of 625ml during the procedure. The bowl was returned to haemonetics for evaluation. Evaluation of the returned bowl confirmed that a leak did occur in the inner core where the bowl base joins with the outer part of the inner core. This could be the possible cause of the leak at the dynamic seal however, it was not possible to confirm due to the condition of the returned sample. There is nothing at this time that would indicate that the death was related to the machine used. It was mentioned by the customer used another cell saver and a new set was used, to continue the procedure, but no further information was provided.
Patient Sequence No: 1, Text Type: N, H10


[176869036] On 16 december 2019: haemonetics was informed by the customer of a patient death the next day after surgery. During a surgery with the cell saver elite processing set emitted an alarm for "liquid in the centrifuge". It was reported there was a loss of 250ml in the bowl and from the bowl reservoir. The leak occurred on cycle 4, with a total volume reinfused of 625ml during the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219343-2020-00001
MDR Report Key9644075
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-29
Date of Report2019-12-16
Date of Event2019-12-12
Date Mfgr Received2019-12-16
Device Manufacturer Date2015-07-10
Date Added to Maude2020-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SHAUN FLANAGAN
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON MA 02110
Manufacturer CountryUS
Manufacturer Postal02110
Manufacturer G1HAEMONETICS CORPORATION
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON MA 02110
Manufacturer CountryUS
Manufacturer Postal Code02110
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELL SAVER ELITE PROCESSING SET (225 ML)
Generic NameCELL SAVER ELITE PROCESSING SET (225 ML)
Product CodeCAC
Date Received2020-01-29
Model NumberCSE-P-225
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHAEMONETICS CORPORATION
Manufacturer Address125 SUMMER STREET BOSTON MA 02110 US 02110


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-01-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.