STEALTHSTATION S8 EM ENT SYSTEM 9735669

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-29 for STEALTHSTATION S8 EM ENT SYSTEM 9735669 manufactured by Medtronic Navigation, Inc.

Event Text Entries

[178595818] Patient information was unavailable from the site. A medtronic representative went to the site to test the equipment. A new emitter was plugged in, but the system continued to say localizer faulted and the fault light was illuminated on the breakout box. The lemo connector pins were examined on both the old and new emitter and none looked to be damaged. The system was rebooted without resolution. The communication cable for the em controller was also reseated without resolution. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[178595819] Medtronic received information regarding a navigation system being used for a procedure. It was reported that during a case the flat emitter was giving a localizer fault error. The site was advised to try unplugging and re-plugging in the flat emitter, however, that did not resolve the issue. It was noted that the site did not have a side emitter and decided to search for an additional flat emitter on one of their other systems. There was no reported impact on the patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1723170-2020-00288
MDR Report Key9644719
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-29
Date of Report2020-03-20
Date of Event2020-01-24
Date Mfgr Received2020-02-26
Device Manufacturer Date2017-09-14
Date Added to Maude2020-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTEALTHSTATION S8 EM ENT SYSTEM
Generic NameEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Product CodePGW
Date Received2020-01-29
Returned To Mfg2020-02-19
Model Number9735669
Catalog Number9735669
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.