S5 DOUBLE HEAD PUMP 10-85-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-01-30 for S5 DOUBLE HEAD PUMP 10-85-00 manufactured by Livanova Deutschland Gmbh.

Event Text Entries

[176949742] There was no patient involvement. (b)(4). A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue. He traced the failure back to the inverter board of the touchscreen, in particular to the coil. Since this type of board is obsolete, the technician replaced the touchscreen and the associated parts. The issue was solved. Subsequent functional verification testing was completed without further issues and the unit was returned to service.
Patient Sequence No: 1, Text Type: N, H10


[176949743] Livanova (b)(4) received a report that during setup a s5 system alarmed. The s5 double head pump was restarted and smoke came out of the device. The pump was swapped with another. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611109-2020-00041
MDR Report Key9646573
Report SourceHEALTH PROFESSIONAL
Date Received2020-01-30
Date of Report2020-01-29
Date of Event2020-01-02
Date Mfgr Received2020-01-02
Device Manufacturer Date2008-02-21
Date Added to Maude2020-01-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactENRICO GRECO
Manufacturer Street14401 W. 65TH WAY
Manufacturer CityARVADA CO 80004
Manufacturer CountryUS
Manufacturer Postal80004
Manufacturer G1LIVANOVA DEUTSCHLAND GMBH
Manufacturer StreetLINDBERGHSTR. 25
Manufacturer CityMUNICH 80309
Manufacturer CountryGM
Manufacturer Postal Code80309
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameS5 DOUBLE HEAD PUMP
Generic NameCONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Product CodeDWB
Date Received2020-01-30
Model Number10-85-00
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIVANOVA DEUTSCHLAND GMBH
Manufacturer AddressLINDBERGHSTR. 25 MUNICH 80309 GM 80309


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-30