[176962148] During numerous endoscopy's, the "endogator single use auxiliary water jet connector" that was connected to olympus gastroenterology endoscopes either broke off completely or began to leak while they were using the water jet. I talked to ambulatory care center (acc) nurses, and they said, between their unit and the operating room, over the last 4 to 6 weeks or so, they have had about a dozen of the endogator water jet connectors either fail or break. When the connectors would fail without breaking off, they would leak between the small seal and a connecting piece of plastic, causing limited water to be used in the cleaning process during the endoscopy. If the connector completely broke off, the procedure would need to be stopped until the broken piece could be replaced.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9646658
MDR Report Key9646658
Date Received2020-01-30
Date of Report2020-01-22
Date of Event2020-01-10
Report Date2020-01-23
Date Reported to FDA2020-01-23
Date Reported to Mfgr2020-01-30
Date Added to Maude2020-01-30
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDOGATOR
Generic NameENDOSCOPIC IRRIGATION/SUCTION SYSTEM
Product CodeOCX
Date Received2020-01-30
Model Number100241
Catalog Number100241
Lot Number440895
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDIVATORS INC.
Manufacturer Address14605 28TH AVENUE NORTH MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-30