[178628156] Other relevant device(s) are: product id: 8253410, serial/lot #: (b)(4), udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[178628157] A healthcare provider (hcp) reported that a nerve monitoring device was not working properly. In mode 1 it does not work and in mode 2, according to the service it gives failure. There was no patient impact.
Patient Sequence No: 1, Text Type: D, B5


[183608719] Analysis for the mainframe found out that the device has been received in good condition, there are no deviations found. Analysis for the interface found out that wave spring washers are worn. The fuse of channel stim 1 is defect. The parts have been replaced. The interface has been successfully tested according to manufacturers specifications. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2020-00065
MDR Report Key9647845
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-30
Date of Report2020-03-05
Date of Event2020-01-07
Date Mfgr Received2020-01-07
Device Manufacturer Date2017-11-30
Date Added to Maude2020-01-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNIM-NEURO? 3.0 MAINFRAME
Generic NameSTIMULATOR, NERVE
Product CodeETN
Date Received2020-01-30
Returned To Mfg2020-02-04
Model Number8253402
Catalog Number8253402
Lot Number214472853
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-30

© 2026 FDA.report