LEAD MODEL 304 304-20

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-30 for LEAD MODEL 304 304-20 manufactured by Cyberonics - Houston.

Event Text Entries

[183965949] During a generator replacement due to battery depletion, the patient presented with high impedance once the new generator was connected to the lead. Pre-op diagnostics could not be performed to determine the impedance, since the patient's generator had reached end of service. The leads were adjusted, the surgeon checked for any visual fractures, took steps to relieve back pressure, made sure the nerve was irrigated and multiple system diagnostic tests were performed but high impedance was still observed, this is captured in mfr. Report #1644487-2020-00142. Generator diagnostics were performed with a test resistor and the impedance was normal, so a generator issue was ruled out. The surgeon then replaced the lead and checked for correct placement on the nerve and high impedance was still observed, this is captured in mfr. Report # 1644487-2020-00141. The lead was then replaced again (captured in this mfr. Report) and high impedance was still observed and the patient was closed. A review of device history records was performed and revealed that the lead passed quality control inspection prior to distribution for the lead that was opened but not used. The surgeon then mentioned that she is unsure of the cause of the high impedance, but a fairly thick layer of fibrous tissue was observed on the patient's nerve. No further surgical intervention has occurred to date. No other relevant information has been received to date.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644487-2020-00140
MDR Report Key9648523
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-30
Date of Report2020-01-30
Date of Event2020-01-08
Date Mfgr Received2020-01-08
Device Manufacturer Date2019-11-13
Date Added to Maude2020-01-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RACHEL KOHN
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal77058
Manufacturer Phone2812287200
Manufacturer G1CYBERONICS - HOUSTON
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal Code77058
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEAD MODEL 304
Generic NameLEAD
Product CodeMUZ
Date Received2020-01-30
Model Number304-20
Lot Number6497
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCYBERONICS - HOUSTON
Manufacturer Address100 CYBERONICS BLVD HOUSTON TX 77058 US 77058


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-30