3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100A 768901

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-01 for 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100A 768901 manufactured by Vyaire Medical Inc..

Event Text Entries

[184315446] (b)(4). Any additional information received from the customer will be included in a follow-up report. Field service representative noted that there is no internal leak. The loud cooling gas noise is due to out of calibration regulator. Regulator was calibrated and verified performance. The unit passed all performance check.
Patient Sequence No: 1, Text Type: N, H10


[184315447] The customer reported that their 3100a has large internal leak. The customer claimed that there is unusual sound coming from internal component of the ventilator. There is no patient involvement associated with this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2020-11303
MDR Report Key9656364
Report SourceUSER FACILITY
Date Received2020-02-01
Date of Report2020-01-03
Date of Event2020-01-03
Date Mfgr Received2020-01-03
Device Manufacturer Date2009-03-16
Date Added to Maude2020-02-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMINDY FABER
Manufacturer Street26125 N. RIVERWOODS BLVD.
Manufacturer CityMETTAWA IL 60045
Manufacturer CountryUS
Manufacturer Postal60045
Manufacturer Phone8727570116
Manufacturer G1VYAIRE MEDICAL INC.
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Generic NameVENTILATOR, HIGH FREQUENCY
Product CodeLSZ
Date Received2020-02-01
Model Number3100A
Catalog Number768901
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL INC.
Manufacturer Address22745 SAVI RANCH PKWY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-01

© 2026 FDA.report