SIDEXIS 4 6618735

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-01 for SIDEXIS 4 6618735 manufactured by Sirona Dental Systems Gmbh.

Event Text Entries

[184708034] The investigation from the r&d department has resulted: sidexis 4 is used in combination with sidicom wls and a central pacs system. Sidicom wls is configurated to send automatically the images to the pacs system after the acquisition. Multiple sidexis 4 users at one windows pc's with individual user accounts logged-in, in parallel. After an examination the user does no logout of his windows session and a different user starts with his additional log-in. Additionally, sidexis 4 is not closed from the previous user. If the above listed conditions are given the issue described will occur. The immediate action is to add an additional instruction into the sidicom wls packages that are shipped. Further capa action is ongoing.
Patient Sequence No: 1, Text Type: N, H10


[184708035] X-ray images were additionally linked to wrong patient record.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9614977-2020-00001
MDR Report Key9656563
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-01
Date of Report2020-01-31
Date Mfgr Received2020-01-16
Date Added to Maude2020-02-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494424
Manufacturer G1SIRONA DENTAL SYSTEMS GMBH
Manufacturer StreetFABRIKSTRASSE 31
Manufacturer CityBENSHEIM, D-64625
Manufacturer CountryGM
Manufacturer Postal CodeD-64625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSIDEXIS 4
Generic NameSYSTEM, IMAGE PROCESSING, RADIOLOGICAL
Product CodeLLZ
Date Received2020-02-01
Catalog Number6618735
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIRONA DENTAL SYSTEMS GMBH
Manufacturer AddressFABRIKSTRASSE 31 BENSHEIM, D-64625 GM D-64625


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-01

© 2026 FDA.report