SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D SDM-00001-3D

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-03 for SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D SDM-00001-3D manufactured by Hologic, Inc.

Event Text Entries

[178638655] It was reported that the gantry will randomly rotate completely around on its own. No injury reported. A field engineer was dispatched to the site and determined the c-arm switch assemblies needed to be replaced. Once this was completed the system was working as intended.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220984-2020-00014
MDR Report Key9661521
Report SourceUSER FACILITY
Date Received2020-02-03
Date of Report2020-01-14
Date of Event2020-01-14
Date Mfgr Received2020-01-14
Device Manufacturer Date2014-02-01
Date Added to Maude2020-02-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN FORNIERI
Manufacturer Street36 & 37 APPLE RIDGE ROAD
Manufacturer CityDANBURY CT 06810
Manufacturer CountryUS
Manufacturer Postal06810
Manufacturer Phone2037318491
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D
Generic NameMAMMOGRAPHY SYSTEM
Product CodeOTE
Date Received2020-02-03
Model NumberSDM-00001-3D
Catalog NumberSDM-00001-3D
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC
Manufacturer Address36 & 37 APPLE RIDGE ROAD DANBURY CT 06810 US 06810


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.