CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-02-03 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[177835891] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[177835892] It was reported that the programmer had difficulty retrieving impedance measurements from the analyser. A higher than expected impedance reading was observed on both leads. The user attempted to come on and off pacing and left pacing on for 20-30 seconds. No patient complications have been reported as a result of this event. The device is still in use.
Patient Sequence No: 1, Text Type: D, B5


[188586055] Product analysis: it was noted that there is a known situation with an ongoing investigation linked to it involving impedance measuring due to a refresh issue. As the impedance measurement requires both atrial and ventricular events it could take up to 4 seconds to display the measured value. If the user continues to tap the field faster than that the update will be delayed as every tap forces the product software analyzer (psa) to take a new measurement. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004593495-2020-00092
MDR Report Key9662138
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-02-03
Date of Report2020-03-05
Date of Event2020-01-10
Date Mfgr Received2020-02-10
Device Manufacturer Date2019-06-21
Date Added to Maude2020-02-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-02-03
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900