MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-02-03 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.
[177835891]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[177835892]
It was reported that the programmer had difficulty retrieving impedance measurements from the analyser. A higher than expected impedance reading was observed on both leads. The user attempted to come on and off pacing and left pacing on for 20-30 seconds. No patient complications have been reported as a result of this event. The device is still in use.
Patient Sequence No: 1, Text Type: D, B5
[188586055]
Product analysis: it was noted that there is a known situation with an ongoing investigation linked to it involving impedance measuring due to a refresh issue. As the impedance measurement requires both atrial and ventricular events it could take up to 4 seconds to display the measured value. If the user continues to tap the field faster than that the update will be delayed as every tap forces the product software analyzer (psa) to take a new measurement. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3004593495-2020-00092 |
MDR Report Key | 9662138 |
Report Source | COMPANY REPRESENTATIVE |
Date Received | 2020-02-03 |
Date of Report | 2020-03-05 |
Date of Event | 2020-01-10 |
Date Mfgr Received | 2020-02-10 |
Device Manufacturer Date | 2019-06-21 |
Date Added to Maude | 2020-02-03 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Street | BAYAN LEPAS FREE INDUSTRIAL ZO |
Manufacturer City | BAYAN LEPAS 11900 |
Manufacturer Country | MY |
Manufacturer Postal Code | 11900 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CARELINK SMARTSYNC BASE |
Generic Name | ANALYZER, PACEMAKER GENERATOR FUNCTION |
Product Code | DTC |
Date Received | 2020-02-03 |
Model Number | 24970A |
Catalog Number | 24970A |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Address | BAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-03 |