LEGEND PNEUMATIC MOTOR PM100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-03 for LEGEND PNEUMATIC MOTOR PM100 manufactured by Mdt Powered Surgical Solutions.

Event Text Entries

[186184748] No conclusion can be drawn. No evaluation was performed, as the device was not returned. If the device is returned in the future, product analysis may be performed. The device user manual warnings section includes instructions to check the device for damage before use. If damage is found, the device should not be used. We will continue to track and trend this complaint type. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[186184749] Repair request initiated for device with the report of not working and poor air flow. It was reported that there was no patient or staff impact but there was prolonged exposure to anesthesia and procedure delay to get another drill. Repair request escalated to product event due to customer indication that this report is a complaint.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1625507-2020-00018
MDR Report Key9662352
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-03
Date of Report2020-03-26
Date of Event2020-01-03
Date Mfgr Received2020-02-28
Device Manufacturer Date2017-08-24
Date Added to Maude2020-02-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street4620 NORTH BEACH STREET
Manufacturer CityFT. WORTH TX 761373291
Manufacturer CountryUS
Manufacturer Postal761373291
Manufacturer Phone7635260594
Manufacturer G1MDT POWERED SURGICAL SOLUTIONS
Manufacturer Street4620 N BEACH ST
Manufacturer CityFORT WORTH TX 76137
Manufacturer CountryUS
Manufacturer Postal Code76137
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEGEND PNEUMATIC MOTOR
Generic NameMOTOR, DRILL, PNEUMATIC
Product CodeHBB
Date Received2020-02-03
Model NumberPM100
Catalog NumberPM100
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMDT POWERED SURGICAL SOLUTIONS
Manufacturer Address4620 N BEACH ST FORT WORTH TX 76137 US 76137


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.