DIGITRAK XT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-02-04 for DIGITRAK XT manufactured by Philips Medical Systems.

Event Text Entries

[177581814] Notes from biomed work order: found device error 602 on display, could not clear it with a battery removal reset. Called philips and discovered this was a global error due to the new year, the 602 error is a firmware error related to the device not correctly interpreting the year 2020 and locking up. Philips is aware of the problem and working to create a resolution for it, when that is completed, i will be notified via e-mail. Team was informed all of the digitrak units are no longer functional until the correction is sent out. Manufacturer response for recorders electronic storage data electrocardiography, digitrak xt (per site reporter) our technician has contact philips per case number [redacted]. No further communication has been sent from philips.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9664395
MDR Report Key9664395
Date Received2020-02-04
Date of Report2020-01-06
Date of Event2020-01-06
Report Date2020-01-06
Date Reported to FDA2020-01-06
Date Reported to Mfgr2020-02-04
Date Added to Maude2020-02-04
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIGITRAK XT
Generic NameELECTROCARDIOGRAPH,AMBULATORY(WITHOUT ANALYSIS)
Product CodeMWJ
Date Received2020-02-04
Device AvailabilityY
Device Age5 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS
Manufacturer Address3000 MINUTEMAN ROAD ANDOVER MA 01810 US 01810


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-04

© 2026 FDA.report