MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-02-04 for GE SAM GAS ANALYZER FGGE3000 manufactured by Marquette Electronics, Inc..
[177837117]
The complaint device was returned for evaluation. Incoming visual inspection found no anomalies. Device evaluation is identified that the unit was overheating due to dust build-up around the fan and the readings were out of tolerance confirming the reported event. The dust build-up was removed. The device was calibrated. The front and rear connectors, and housing was checked for damage. The root cause for the reported event is determined to be a result of normal use. This type of event will continue to be monitored. It should be noted that this is being filed based on retrospective review.
Patient Sequence No: 1, Text Type: N, H10
[177837118]
Reportedly, post repair, the readings were incorrect. There was no report of patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3007409280-2020-00029 |
MDR Report Key | 9665815 |
Report Source | CONSUMER,HEALTH PROFESSIONAL |
Date Received | 2020-02-04 |
Date of Report | 2019-02-04 |
Date Mfgr Received | 2019-02-04 |
Date Added to Maude | 2020-02-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | JENNA VARGAS |
Manufacturer Street | 212 AVENIDA FABRICANTE |
Manufacturer City | SAN CLEMENTE CA 92672 |
Manufacturer Country | US |
Manufacturer Postal | 92672 |
Manufacturer Phone | 8004495328 |
Manufacturer G1 | AVANTE HEALTH SOLUTIONS |
Manufacturer Street | 212 AVENIDA FABRICANTE |
Manufacturer City | SAN CLEMENTE CA 92672 |
Manufacturer Country | US |
Manufacturer Postal Code | 92672 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | GE SAM GAS ANALYZER |
Generic Name | SMART ANESTHESIA MULTI-GAS (SAM) MODULE |
Product Code | CCK |
Date Received | 2020-02-04 |
Returned To Mfg | 2019-02-07 |
Model Number | FGGE3000 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MARQUETTE ELECTRONICS, INC. |
Manufacturer Address | 8200 WEST TOWER AVE. MILWAUKEE WI 53223 US 53223 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-04 |