GE SAM GAS ANALYZER FGGE3000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-02-04 for GE SAM GAS ANALYZER FGGE3000 manufactured by Marquette Electronics, Inc..

Event Text Entries

[177837117] The complaint device was returned for evaluation. Incoming visual inspection found no anomalies. Device evaluation is identified that the unit was overheating due to dust build-up around the fan and the readings were out of tolerance confirming the reported event. The dust build-up was removed. The device was calibrated. The front and rear connectors, and housing was checked for damage. The root cause for the reported event is determined to be a result of normal use. This type of event will continue to be monitored. It should be noted that this is being filed based on retrospective review.
Patient Sequence No: 1, Text Type: N, H10


[177837118] Reportedly, post repair, the readings were incorrect. There was no report of patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007409280-2020-00029
MDR Report Key9665815
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2020-02-04
Date of Report2019-02-04
Date Mfgr Received2019-02-04
Date Added to Maude2020-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNA VARGAS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal92672
Manufacturer Phone8004495328
Manufacturer G1AVANTE HEALTH SOLUTIONS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal Code92672
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGE SAM GAS ANALYZER
Generic NameSMART ANESTHESIA MULTI-GAS (SAM) MODULE
Product CodeCCK
Date Received2020-02-04
Returned To Mfg2019-02-07
Model NumberFGGE3000
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMARQUETTE ELECTRONICS, INC.
Manufacturer Address8200 WEST TOWER AVE. MILWAUKEE WI 53223 US 53223


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.