ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER N/A ULT8.5-38-40-P-32S-CLB-RH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-02-04 for ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER N/A ULT8.5-38-40-P-32S-CLB-RH manufactured by Cook Inc.

Event Text Entries

[177843695] Occupation: unknown. Initial reporter also sent report to fda: unknown. Pma/510(k) #: preamendment. Investigation? Evaluation. A review of the complaint history, device history record, drawing, instructions for use (ifu), quality control, and specifications of the device, as well as a visual inspection, functional test, and dimensional verification, were conducted during the investigation. The complainant returned on catheter and flexible stiffener for investigation. Physical examination of the returned device showed the flexible stiffener inserted in the catheter and biomatter on the exterior of the device. The flexible stiffener was found to be advanced all the way to the distal tip of the catheter, but was extending from the mac-loc hub. The flexible stiffener was able to be removed by relaxing the distal tip of the catheter, and found multiple sections of elongation. All dimensions deemed relevant to the reported failure were analyzed (flexible stiffener od, catheter id), and confirmed that the device was manufactured within specification. Additionally, a document-based investigation evaluation was performed. Sufficient inspection activities are in place to identify this failure mode prior to distribution. The device history record (dhr) for the complaint lot, catheter tubing subassemblies, and flexible stiffener raw material lots revealed no related nonconformances. A database search for complaints on the reported lot found no additional complaints reported from the field. At this time, there is no evidence suggesting that nonconforming product from this lot exists in house or in the field. Product labeling was also reviewed. The product instructions for use (ifu) provides the following information to the user related to the reported failure mode: precautions:? When inserting a stiffening cannula into a catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of suture.? How supplied:? Supplied sterilized by ethylene oxide gas in peel-open packages. Intended for one-time use. Sterile if package is unopened or undamaged. Do not upset he product if there is doubt as to whether the product is sterile. Store in a dark, dry, cool place. Avoid extended exposure to light. Upon removal from package, inspect the product to ensure no damage has occurred.? Based on the information provided, visual inspection of the returned product, and the results of the document investigation, cook has concluded that a component failure without a manufacturing or design defect contributed to this incident. The appropriate personnel have been notified. Per the quality engineering risk assessment no further action is required. Cook will continue to monitor for similar complaints.
Patient Sequence No: 1, Text Type: N, H10


[177843696] As reported, an unknown patient required the placement of an ultrathane mac-loc locking loop biliary drainage catheter for an unknown procedure. The operator reported that during the procedure, "they could not get the stylet to progress through the catheter and it would get stuck. " the same failure was noted with another device of the same lot. The device was returned with the flexible stiffener advanced in the catheter, and could only be removed by relaxing the distal tip of the catheter. Examination of returned product also noted multiple sections of elongation of the flexible stiffener. As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00284
MDR Report Key9667263
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-02-04
Date of Report2020-01-31
Date of Event2019-10-31
Date Mfgr Received2020-01-15
Device Manufacturer Date2019-07-15
Date Added to Maude2020-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUNK JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER
Generic NameGCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY
Product CodeGCA
Date Received2020-02-04
Returned To Mfg2020-01-13
Model NumberN/A
Catalog NumberULT8.5-38-40-P-32S-CLB-RH
Lot Number9879095
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-04