PHILIPS M1019A G5 ANESTHETIC GAS ANALYZER FGPH1019

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-02-04 for PHILIPS M1019A G5 ANESTHETIC GAS ANALYZER FGPH1019 manufactured by Draeger Medical, Inc..

Event Text Entries

[177837280] The complaint device was returned for evaluation. Incoming visual inspection found that the warranty label was compromised. Device evaluation is identified that the dir sensor was causing bad co2 readings confirming the reported event. The dir sensor was replaced and the device was calibrated testing to oem specifications. The case was also checked for damage. The root cause for the reported event is determined to be the aged dir sensor. This type of event will continue to be monitored. It should be noted that this is being filed based on retrospective review.
Patient Sequence No: 1, Text Type: N, H10


[177837281] Reportedly, post repair, the co2 keeps climbing when connected to a patient. There was no report of patient harm. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007409280-2020-00030
MDR Report Key9667355
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2020-02-04
Date of Report2018-12-26
Date Mfgr Received2018-12-26
Date Added to Maude2020-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNA VARGAS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal92672
Manufacturer Phone8004495328
Manufacturer G1AVANTE HEALTH SOLUTIONS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal Code92672
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePHILIPS M1019A G5 ANESTHETIC GAS ANALYZER
Generic NameANESTHETIC GAS MONITOR
Product CodeCCK
Date Received2020-02-04
Returned To Mfg2019-01-08
Model NumberFGPH1019
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDRAEGER MEDICAL, INC.
Manufacturer Address3135 QUARRY RD. TELFORD PA 18969 US 18969


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.