MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-02-04 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.
[177940460]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[177940461]
It was reported that during an implant procedure the p waves being measured through the analyzer did not match the p waves measured through the implantable heart device. The p waves measured greater through the analyzer than through the implantable device, even at the most sensitive value programmed. It was further reported that the programmer was displaying diagnostic messages indicating that the bluetooth connection was lost. The measurements were tried again but with the same result and again the diagnostic message of "bluetooth connection lost" displayed. The user switched the implantable device mode from ddd to aai and the measurements through the device then matched the ones from the analyzer measurement and the "connection lost" message no longer displayed. It was noted by the user that they were only two or three feet from the implanted device and at one point was "right on top of the device" when getting the connection lost messages. The issue was escalated to the programmer development team. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
[184375132]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3004593495-2020-00097 |
MDR Report Key | 9667864 |
Report Source | COMPANY REPRESENTATIVE |
Date Received | 2020-02-04 |
Date of Report | 2020-03-02 |
Date of Event | 2020-01-08 |
Date Mfgr Received | 2020-02-19 |
Device Manufacturer Date | 2018-12-07 |
Date Added to Maude | 2020-02-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Street | BAYAN LEPAS FREE INDUSTRIAL ZO |
Manufacturer City | BAYAN LEPAS 11900 |
Manufacturer Country | MY |
Manufacturer Postal Code | 11900 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CARELINK SMARTSYNC BASE |
Generic Name | ANALYZER, PACEMAKER GENERATOR FUNCTION |
Product Code | DTC |
Date Received | 2020-02-04 |
Model Number | 24970A |
Catalog Number | 24970A |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Address | BAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-04 |