CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-02-04 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[177940460] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[177940461] It was reported that during an implant procedure the p waves being measured through the analyzer did not match the p waves measured through the implantable heart device. The p waves measured greater through the analyzer than through the implantable device, even at the most sensitive value programmed. It was further reported that the programmer was displaying diagnostic messages indicating that the bluetooth connection was lost. The measurements were tried again but with the same result and again the diagnostic message of "bluetooth connection lost" displayed. The user switched the implantable device mode from ddd to aai and the measurements through the device then matched the ones from the analyzer measurement and the "connection lost" message no longer displayed. It was noted by the user that they were only two or three feet from the implanted device and at one point was "right on top of the device" when getting the connection lost messages. The issue was escalated to the programmer development team. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[184375132] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004593495-2020-00097
MDR Report Key9667864
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-02-04
Date of Report2020-03-02
Date of Event2020-01-08
Date Mfgr Received2020-02-19
Device Manufacturer Date2018-12-07
Date Added to Maude2020-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-02-04
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-04