MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-04 for AXIOS M00553640 5364 manufactured by Boston Scientific Corporation.
[178129121]
(b)(4). The device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10
[178129122]
It was reported to boston scientific corporation that a hot axios stent was to be implanted transduodenally to a pseudocyst during a drainage procedure performed on (b)(6) 2020. According to the complainant, during the procedure, the first flange of the axios was deployed inside the cyst, however, the flange did not fully expand. The second flange was deployed, the first flange then expanded but the stent had come out of the cyst and it was no longer in the correct position. The stent was removed using forceps and the procedure was completed using a 15mm axios. There were no patient complications reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005099803-2020-00229 |
MDR Report Key | 9668027 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-02-04 |
Date of Report | 2020-02-04 |
Date of Event | 2020-01-10 |
Date Mfgr Received | 2020-01-12 |
Device Manufacturer Date | 2019-05-17 |
Date Added to Maude | 2020-02-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | CAROLE MORLEY |
Manufacturer Street | 300 BOSTON SCIENTIFIC WAY |
Manufacturer City | MARLBOROUGH MA 01752 |
Manufacturer Country | US |
Manufacturer Postal | 01752 |
Manufacturer Phone | 5086834015 |
Manufacturer G1 | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Street | BALLYBRIT BUSINESS PARK |
Manufacturer City | GALWAY |
Manufacturer Country | EI |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AXIOS |
Generic Name | PANCREATIC STENT, COVERED, METALLIC, REMOVABLE |
Product Code | PCU |
Date Received | 2020-02-04 |
Model Number | M00553640 |
Catalog Number | 5364 |
Lot Number | 0023808088 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Address | 300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-02-04 |