EVOTECH 50014

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-02-05 for EVOTECH 50014 manufactured by Advanced Sterilization Products Services Inc..

Event Text Entries

[177800049] Brand new evotech installed. The spray arm has not been functioning correctly causing the cycles to fail. The issue was intermittent, and the asp technicians have been out several times and have tried several different solutions. Eventually, they told us that it was a manufacturer's defect and they replaced the part altogether. Device is evotech scope washer, catalog #50014, gtin: (b)(4). Manufacturer response for accessories, cleaning, for endoscope, (evotech) (per site reporter). They have offered to trade it out or give us a two-year warranty.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9670500
MDR Report Key9670500
Date Received2020-02-05
Date of Report2020-01-27
Date of Event2019-12-21
Report Date2020-02-03
Date Reported to FDA2020-02-03
Date Reported to Mfgr2020-02-05
Date Added to Maude2020-02-05
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVOTECH
Generic NameACCESSORIES, CLEANING, FOR ENDOSCOPE
Product CodeFEB
Date Received2020-02-05
Model Number50014
Catalog Number50014
Device AvailabilityY
Device Age1 DA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS SERVICES INC.
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-05

© 2024 FDA.report