PILLCAM RECORDER DR3 FGS-0347

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-02-05 for PILLCAM RECORDER DR3 FGS-0347 manufactured by Given Imaging, Inc..

Event Text Entries

[177818615] Inpatient in ticu was scheduled for capsule endoscopy. Protocol followed to prepare and pair capsule and recorder without any difficulties. Nurse had patient swallow capsule. Within 10 minutes, recorder flashing yellow light every 5 seconds indicating that the recorder was not receiving signals from the capsule. Nurse called medtronic tech support to troubleshoot problem. After attempts to correct issue, the recorder did continue with the flashing yellow light every 5 seconds. When the recorder was downloaded, the study showed the capsule signals were not detected by the recorder. The study was incomplete. The conclusion was that the patient was on telemetry which can cause interference with the capsule study. Using different equipment, the capsule study was repeated with the patient off telemetry (patient remained in ticu). There were episodes where the yellow light would flash every 5 seconds. The study did show that there was interference and that the capsule signal was not received by the recorder. This study was also incomplete. Pillcam recorder ref: fgs-0347 and capsule. Capsule ids: (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9671047
MDR Report Key9671047
Date Received2020-02-05
Date of Report2020-01-09
Date of Event2020-01-07
Report Date2020-01-09
Date Reported to FDA2020-01-09
Date Reported to Mfgr2020-02-05
Date Added to Maude2020-02-05
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePILLCAM RECORDER DR3
Generic NameSYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, CAPSULE
Product CodeNEZ
Date Received2020-02-05
Catalog NumberFGS-0347
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerGIVEN IMAGING, INC.
Manufacturer Address15 HAMPSHIRE STREET MANSFIELD MA 02048 US 02048


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-05