SIMPLYGO MINI 1113602

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-05 for SIMPLYGO MINI 1113602 manufactured by Philips Respironics Inc..

Event Text Entries

[177790688] The manufacturer received information alleging an oxygen concentrator that shut down during patient use. The patient was admitted to the hospital. The device has yet to be returned to the manufacturer for evaluation. A follow up report will be submitted when the manufacturer has completed the investigation.
Patient Sequence No: 1, Text Type: D, B5


[184085203] The manufacturer previously reported an oxygen concentrator that allegedly shut down during patient use. The patient was admitted to the hospital. The device was returned to the manufacturer's product investigation laboratory for further evaluation. The device's downloaded event log was reviewed and was found to have no errors related to the event. The device was tested and was found to operate and alarm to design specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2518422-2020-00292
MDR Report Key9671072
Report SourceUSER FACILITY
Date Received2020-02-05
Date of Report2020-01-31
Date of Event2020-01-18
Date Mfgr Received2020-02-12
Device Manufacturer Date2018-03-22
Date Added to Maude2020-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactADAM PRICE
Manufacturer Street1001 MURRY RIDGE LANE
Manufacturer CityMURRYSVILLE PA 15668
Manufacturer CountryUS
Manufacturer Postal15668
Manufacturer Phone7243349303
Manufacturer G1PHILIPS RESPIRONICS INC.
Manufacturer Street1001 MURRY RIDGE LANE
Manufacturer CityMURRYSVILLE PA 15668
Manufacturer CountryUS
Manufacturer Postal Code15668
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSIMPLYGO MINI
Generic NameGENERATOR, OXYGEN, PORTABLE
Product CodeCAW
Date Received2020-02-05
Returned To Mfg2020-02-07
Model Number1113602
Catalog Number1113602
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS RESPIRONICS INC.
Manufacturer Address1001 MURRY RIDGE LANE MURRYSVILLE PA 15668 US 15668


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-05