CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-02-05 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[177854659] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[177854660] It was reported that the programmer displayed a remaining longevity of four years for the cardiac resynchronisation therapy pacemaker ( crt-p). When a different programmer was used the remaining longevity was displayed as one month and an earlier remote monitoring network transmission displayed a remaining longevity of two months. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[180462935] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2182208-2020-00231
MDR Report Key9672533
Report SourceHEALTH PROFESSIONAL
Date Received2020-02-05
Date of Report2020-02-26
Date of Event2020-01-22
Date Mfgr Received2020-02-20
Device Manufacturer Date2018-07-02
Date Added to Maude2020-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS MN 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-02-05
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS MN 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-05