IFUSE IMPLANT SYSTEM UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-02-05 for IFUSE IMPLANT SYSTEM UNKNOWN manufactured by Si-bone, Inc..

Event Text Entries

[180556560] Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmea, the most probable root cause is user error; initial malpositioning of the implant and/or improper implant size selection.
Patient Sequence No: 1, Text Type: N, H10


[180556561] The patient had right side si joint arthrodesis in (b)(6) 2019 where two implants were installed. The patient later reported to a new surgeon with radicular pain complaints. The implants were slightly malpositioned and the patient wanted them removed. In (b)(6) 2020, the surgeon performed a revision procedure where he removed both implants using chisels as they were solidly fixed in bone. The status of the patient following the revision procedure is not known.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007700286-2020-00004
MDR Report Key9673201
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-02-05
Date of Report2020-02-05
Date of Event2020-01-15
Date Mfgr Received2020-01-15
Date Added to Maude2020-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactW. RECKLING, M.D.
Manufacturer Street471 EL CAMINO REAL SUITE 101
Manufacturer CitySANTA CLARA CA 950504482
Manufacturer CountryUS
Manufacturer Postal950504482
Manufacturer Phone4082070700
Manufacturer G1SI-BONE, INC.
Manufacturer Street471 EL CAMINO REAL SUITE 101
Manufacturer CitySANTA CLARA CA 950504482
Manufacturer CountryUS
Manufacturer Postal Code950504482
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIFUSE IMPLANT SYSTEM
Generic NameORTHOPEDIC ROD
Product CodeOUR
Date Received2020-02-05
Model NumberUNKNOWN
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSI-BONE, INC.
Manufacturer Address471 EL CAMINO REAL SUITE 101 SANTA CLARA CA 950504482 US 950504482


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-05