CENTRIMAG 2ND GENERATION PRIMARY CONSOLE 201-30300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-05 for CENTRIMAG 2ND GENERATION PRIMARY CONSOLE 201-30300 manufactured by Thoratec Switzerland Gmbh.

Event Text Entries

[179790317] The device is expected to be returned for evaluation. It has not yet been received. The centrimag motor associated with this event is being reported under mfr # 2916596-2020-00551. No further information was provided. A supplemental report will be submitted when the manufacturer? S investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[179790318] It was reported that there was rapid motor deterioration, and a whirring sound was emitted from the device. The console screen then went black and the motor failed. After a few seconds the console rebooted but the decision was made to exchange the entire unit. When the pump head was removed it was noted that the motor was extremely hot. The pump was inserted into a new motor and console. During this exchange the patient experienced mild desaturation but there was no haemodynamic compromise due to the quick exchange of the pump. There were no alarms reported for the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916596-2020-00550
MDR Report Key9673915
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-05
Date of Report2020-02-05
Date of Event2019-12-08
Date Mfgr Received2020-01-15
Date Added to Maude2020-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BOB FRYC
Manufacturer Street6035 STONERIDGE DRIVE
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal94588
Manufacturer Phone7818528204
Manufacturer G1THORATEC SWITZERLAND GMBH
Manufacturer StreetTECHNOPARKSTRASSE 1
Manufacturer CityZURICH CH-8005
Manufacturer CountrySZ
Manufacturer Postal CodeCH-8005
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Generic NameCONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Product CodeDWA
Date Received2020-02-05
Model Number201-30300
Catalog Number201-30300
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTHORATEC SWITZERLAND GMBH
Manufacturer AddressTECHNOPARKSTRASSE 1 ZURICH CH-8005 SZ CH-8005


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.