MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-06 for ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER N/A ULT8.5-38-25-P-6S-CLM-RH manufactured by Cook Inc.
[178142835]
Date of event: unknown. Pma/510(k) #: exempt. This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10
[178142836]
It was reported an unknown patient required placement of an ultrathane mac-loc locking loop multipurpose drainage catheter for an unknown procedure. During the procedure, the operator reported the "drain material" was "so damaged by concertina-ing". This device was removed due to the damage and replaced with a new device. The operator placed the second device with the "stiffener half pulled back" and did not come out "as easily as is usually expected". As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1820334-2020-00298 |
MDR Report Key | 9676949 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-02-06 |
Date of Report | 2020-04-03 |
Date Mfgr Received | 2020-04-03 |
Date Added to Maude | 2020-02-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. JENNIFER CANADA |
Manufacturer Street | 750 DANIELS WAY |
Manufacturer City | BLOOMINGTON IN 47404 |
Manufacturer Country | US |
Manufacturer Postal | 47404 |
Manufacturer Phone | 8123392235 |
Manufacturer G1 | COOK INC. |
Manufacturer Street | 750 DANIELS WAY |
Manufacturer City | BLOOMINGTON IN 47404 |
Manufacturer Country | US |
Manufacturer Postal Code | 47404 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER |
Generic Name | GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY |
Product Code | GBO |
Date Received | 2020-02-06 |
Model Number | N/A |
Catalog Number | ULT8.5-38-25-P-6S-CLM-RH |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COOK INC |
Manufacturer Address | 750 DANIELS WAY BLOOMINGTON IN 47404 US 47404 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-06 |