[178142034] Pma/510k #: k170622. This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[178142035] It was reported that during a bakri balloon insertion for management of a postpartum hemorrhage due to cervical laceration and uterine atony, the syringe of the cook bakri postpartum balloon with rapid instillation components broke off in the hub preventing inflation. Around four hours after delivery, the patient had lost 2500 ml of blood and the first cook bakri balloon was placed transvaginally. After it was placed the balloon was being filled with saline when the syringe broke off within the stopcock. Another bakri balloon was used to achieve hemostasis. The patient received four units of prbc after losing 1500 ml of blood. No adverse effects related to the alleged malfunction were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00300
MDR Report Key9677782
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-02-06
Date of Report2020-03-13
Date of Event2020-02-04
Date Mfgr Received2020-03-09
Device Manufacturer Date2019-11-20
Date Added to Maude2020-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS
Generic NameOQY INTRAUTERINE BALLOON
Product CodeOQY
Date Received2020-02-06
Returned To Mfg2020-02-13
Model NumberG24237
Catalog NumberJ-SOSR-100500
Lot Number10162713
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-06