PRIM CONS 1ST GEN, STANDALONE 201-30100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-06 for PRIM CONS 1ST GEN, STANDALONE 201-30100 manufactured by Thoratec Switzerland Gmbh.

Event Text Entries

[178097525] The motor associated with this event is being reported under mfr # 2916596-2020-00219. No further information was provided. A supplemental report will be submitted when the manufacturer? S investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[178097526] It was reported that the patient was implanted with centrimag and a sudden "flow below minimum" alarm occurred without any manipulation of the patient or system. At the same time was noted on the rpm display. The clinical team noted that there was no blood flow through the pump and the pump had stopped. The primary console, motor, and flow probe were immediately changed and blood flow was restored. The patient had adequate oxygenation and was hemodynamically stable during the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916596-2020-00705
MDR Report Key9678356
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-06
Date of Report2020-02-06
Date of Event2020-01-11
Date Mfgr Received2020-01-13
Date Added to Maude2020-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BOB FRYC
Manufacturer Street6035 STONERIDGE DRIVE
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal94588
Manufacturer Phone7818528204
Manufacturer G1THORATEC SWITZERLAND GMBH
Manufacturer StreetTECHNOPARKSTRASSE 1
Manufacturer CityZURICH CH-8005
Manufacturer CountrySZ
Manufacturer Postal CodeCH-8005
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRIM CONS 1ST GEN, STANDALONE
Generic NameCONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Product CodeDWA
Date Received2020-02-06
Returned To Mfg2020-01-24
Model Number201-30100
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTHORATEC SWITZERLAND GMBH
Manufacturer AddressTECHNOPARKSTRASSE 1 ZURICH CH-8005 SZ CH-8005


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2020-02-06