MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-06 for PROTAPER NEXT NITIFILE X1 25MM A080022500103 manufactured by Dentsply Maillefer.
[178618845]
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files. This event, therefore, is reportable per 21cfr part 803. The returned protaper next nitifile x1 25mm is actually broken in the active part (fatigue). No material defect was found during analysis of the rupture pattern. No unused file is available for evaluation. Nothing unusual to report was found during dhr review (batch #1579983). Root causes are not identified. We will track this kind of event and monitor the trend.
Patient Sequence No: 1, Text Type: N, H10
[178618846]
In this event it was reported that a protaper next broke during use. The separated piece was still in the tooth as this mdr evaluation.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 8031010-2020-00006 |
MDR Report Key | 9679861 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2020-02-06 |
Date of Report | 2020-02-06 |
Date Mfgr Received | 2020-01-07 |
Date Added to Maude | 2020-02-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | DENTIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. KARL NITTINGER |
Manufacturer Street | 221 W. PHILADELPHIA ST. SUITE 60W |
Manufacturer City | YORK PA 17401 |
Manufacturer Country | US |
Manufacturer Postal | 17401 |
Manufacturer Phone | 7178494229 |
Manufacturer G1 | DENTSPLY MAILLEFER |
Manufacturer Street | CHEMIN DU VERGER 3 |
Manufacturer City | BALLAIGUES, 1338 |
Manufacturer Country | SZ |
Manufacturer Postal Code | 1338 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PROTAPER NEXT NITIFILE X1 25MM |
Generic Name | FILE, PULP CANAL, ENDODONTIC |
Product Code | EKS |
Date Received | 2020-02-06 |
Returned To Mfg | 2020-01-07 |
Model Number | NA |
Catalog Number | A080022500103 |
Lot Number | 1579983 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | DENTSPLY MAILLEFER |
Manufacturer Address | CHEMIN DU VERGER 3 BALLAIGUES, 1338 SZ 1338 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-06 |