CHEMOLOCK BAG SPIKE - A CSTD COMPONENT CL-3955

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-02-06 for CHEMOLOCK BAG SPIKE - A CSTD COMPONENT CL-3955 manufactured by Icu Medical, Inc..

Event Text Entries

[178348761] Cstd dropped out of chemotherapy infusion bag as the nurse was hanging the bag to start infusion. Both the patient and nurse were exposed. The infusion of chemotherapy (carboplatin) was closed using the icu medical cstd, chemolock until the cl-3955 fell out of the bag resulting in approximately 50ml of hazardous solution falling on the nurse, patient, patient? S chair and the floor where the nurse was standing. The nurse was able to insert the spike back into the bag to stop flow. The bag/tubing was contained in chemo zip lock bag then returned to the pharmacy for black bucket bulk chemo waste disposal. The pharmacy compounded a second bag for treatment of the patient. After a spill during training ((b)(6) 2019) the icu medical representatives supplied clips prevent bag spike from falling out of bbraun non-pvc bags and later a second bag clip for the hospira/icu medical button type of infusion bags. They did not mention a 3rd type of bag clip for the type of bag involved in our spill after the spill occurred, they could have prevented this exposure/spill if they had let us know. We use the 1st and 2nd type of bag clips routinely on those bags but we had no knowledge of the possibly of icu medical ndc 0990-7984-37 failing to retain the chemolock cl-3955 bag spike. Fda safety report id # (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5092828
MDR Report Key9681767
Date Received2020-02-06
Date of Report2020-02-04
Date of Event2020-02-03
Date Added to Maude2020-02-07
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHEMOLOCK BAG SPIKE - A CSTD COMPONENT
Generic NameCLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTRUCTION AND TRANSFER SYSTEM
Product CodeONB
Date Received2020-02-06
Model NumberCL-3955
Lot Number4344340
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerICU MEDICAL, INC.


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-06